VGSM Technology (M)
Sdn Bhd
Sdn Bhd
No.34G, 34A & 34B, Lorong Perda Selatan 1, Bandar Perda, 14000 Bukit Mertajam, Penang, Malaysia.
+604-538 1998
+604-530 1998
+604-530 1998
PH10T
Previous | 5 / 9 | Next |
PH10T (Fully Motorised Probe Head)
The PH10T is a fully motorised probe head that offers immediate indexing from 0° to 105° in the A axis and through 360° in the B axis to be moved to 720 repeatable positions in 7.5 degree increments. This probe head should be used by customers requiring frequent indexing or when more precise alignment to the features being measured is required.
TP20 Stylus Modules
A range of seven, application specific, probe modules is available and are identified by coloured caps.
Change Rack MCR20
The MCR20 probe module changing rack is designed to securely hold stored modules for rapid automatic changing, and to protect them from airborne contaminants which may be present within the working environment.
TP200 Stylus Modules
The stylus module is mounted on the TP200 / TP200B probe body via a highly repeatable magnetic kinematic joint, providing a rapid stylus changing capability and probe overtravel protection.
There are three modules available, with two different overtravel forces.
The SCR200 provides automatic, high speed changing of up to six TP200 stylus modules. The SCR200 is powered by the separate probe interface, PI 200-3, and provides features to facilitate safe stylus changing. The SCR200 is available in low force and standard force kits, where each kit comprises one SCR200 plus three stylus modules of the same force.
The PH10T is a fully motorised probe head that offers immediate indexing from 0° to 105° in the A axis and through 360° in the B axis to be moved to 720 repeatable positions in 7.5 degree increments. This probe head should be used by customers requiring frequent indexing or when more precise alignment to the features being measured is required.
Length | 103 mm |
Width | 62 mm |
Weight | 645 g |
Probe mounting | M8 thread (2-wire) |
Head mounting | Shank to suit CMM |
Suitable controller | PHC10-3 PLUS |
Repeatability of position | 0.4 µm (0.00002 in) specified at a distance of 100 mm |
Total number of positions | 720 |
Maximum recommended drive torque | 0.45 Nm |
Maximum extension bar | 300 mm (11.8 in) using PEL4 extension |
TP20 Stylus Modules
A range of seven, application specific, probe modules is available and are identified by coloured caps.
Module | Application | Trigger forces* | Stylus lengths | |
---|---|---|---|---|
SF- Standard force |
|
Suited to the majority of applications |
XY: 0.08 N Z: 0.75 N Stylus: 10 mm |
10 mm to 50 mm GF |
LF- Low force |
|
Low trigger forces applications - e.g. rubber seals |
XY: 0.055 N Z: 0.65 N Stylus: 10 mm |
10 mm to 30 mm |
MF- Medium force |
|
Where a higher trigger force than standard is required |
XY: 0.1 N Z: 1.9 N Stylus: 25 mm |
10 mm to 60 mm |
EF- Extended force |
|
Large stylus assemblies or where vibration causes spurious 'air' triggers |
XY: 0.1 N Z: 3.2N Stylus: 50 mm |
10 mm to 60 mm |
6W- 6-way |
|
Measurement in the –Z direction - e.g. undercuts |
XY: 0.14 N Z: 1.6N Stylus: 10 mm |
10 mm to 30 mm |
EM1 STD- 50 mm extension |
|
Majority of applications requiring a 50 mm extension |
XY: 0.08 N Z: 0.75 N Stylus: 10 mm |
10 mm to 50 mm |
EM2 STD- 75 mm extension |
|
Majority of applications requiring a 75 mm extension |
XY: 0.08 N Z: 0.75 N Stylus: 10 mm |
10 mm to 50 mm |
Change Rack MCR20
The MCR20 probe module changing rack is designed to securely hold stored modules for rapid automatic changing, and to protect them from airborne contaminants which may be present within the working environment.
TP200 Stylus Modules
The stylus module is mounted on the TP200 / TP200B probe body via a highly repeatable magnetic kinematic joint, providing a rapid stylus changing capability and probe overtravel protection.
There are three modules available, with two different overtravel forces.
Module | Application | Comments |
SF (standard force) | General use. | Styli up to 100 mm long and ball diameter > 1 mm. |
LF (low force) | Small diameter stylus balls or where minimum force is essential. | Balls less than 1 mm diameter. |
EO (extended overtravel) | Extra overtravel to enable the CMM to stop safely from higher probing speeds. |
Same overtravel force as SF. Extra 8 mm of overtravel in probe Z-axis. |
SCR200 Change Rack
The SCR200 provides automatic, high speed changing of up to six TP200 stylus modules. The SCR200 is powered by the separate probe interface, PI 200-3, and provides features to facilitate safe stylus changing. The SCR200 is available in low force and standard force kits, where each kit comprises one SCR200 plus three stylus modules of the same force.